User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: HP 6038A:System Power Supply DATE: 17-Nov-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 46 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ X 1.002 ASK- R N P F W 1.003 HEAD INITIAL CONDITIONS 1.004 DISP The following test equipment is required: 1.004 DISP 1. Metered vaiac DTB # 2-172 1.004 DISP 2. Transistor Devices DLP 50-150-3000A 1.004 DISP 3. Sensitive Research 20 Amp Shunt DTB # 14-109 1.005 DISP Set the UUT front panel controls as follows: 1.005 DISP LINE ON/OFF ........................................ OFF 1.005 DISP OUTPUT ADJUST ................................ fully CCW 1.006 DISP Connect the System port 1 to the UUT HP-IB connector. 1.007 DISP Remove the UUT rear panel terminal strip cover. 1.008 DISP All connections are made to the rear panel 1.008 DISP terminal strip. 1.009 DISP Connect the UUT and the 1281 as follows: 1.009 DISP 6038A TO 1281 1.009 DISP OUTPUT (+S) ............> INPUT-HIGH 1.009 DISP OUTPUT (-S) ............> INPUT-LO 1.010 DISP Connect UUT AC line cord to Metered Variac. 1.011 DISP Set the variac to 115 VAC. 1.012 DISP Set the UUT front panel controls as follows: 1.012 DISP LINE ON/OFF ......................................... ON 1.013 IEEE [D3000] 1.014 CALL HP 6038A: Sub 1 1.015 CALL HP 6038A: Sub 2 1.016 IEEE CLR; 1.017 HEAD 1.018 DISP Set the UUT front panel controls as follows: 1.018 DISP LINE ON/OFF ........................................ OFF 1.019 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 18 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.